ExOsense Piezo-Resonant Liquid Level
| |
|
Quick Specs...
- Non-intrusive, never touches the fluid
- Eliminates fluid contamination
- No calibration needed
- Simple installation, no special mounting
- Eliminates fluid compatibility issues
- Works on most plastic bottle materials
|
|
|
|
|
| |
ExOsense Wins Bronze in 12th Annual Engineering Awards
|
|
|
Overview / Design Principle:
The new ExOsense is the first and only affordable, non-intrusive liquid level sensor for plastic fluid containers. ExOsense sensors adhere to the outside of plastic bottles and are unaffected by the color or transparency of the plastic. Liquids inside the bottle are untouched, so with ExOsense there is no issue of material compatibility or contamination. Best of all, ExOsense sensors fit any size and shape vessel, from small bottles, to large tanks.
The Exosense sensor head features a peel-and-stick adhesive face that can be affixed anywhere on the outside of the tank to provide high, low or any intermediate point level fluid sensing. A detachable USB type mini-cable feeds the solid-state switch unit.
Equipment Used On: Hematology, Immuno-chemistry, Histology, Medical Laser Systems, Hemodialysis, Cytology.
Applications:
- Fluid Monitoring Application
- Reagents
- Waste
- Diluent
- Detergent/Wash
- Dialysate
- Coolant
- Sterile Water
|
|
| Specifications: |
Compatible Plastic Bottle Materials
|
Polyethylene (PE), Polypropylene (PP), Polysufone (PS), Polycarbonate (PC) , ABS, Styrene, PVC, and others
|
Bottle Materials Not Recommended
|
Teflon© family, or any foamed core plastic
|
Min. Bottle Diameter
|
3" (76.2 mm)
|
Bottle Wall Thickness
|
0.04" to 0.15" (1.00mm to 3.8mm)
|
Termination of Transducer
|
Mini USB Connector to Electronics Control Module (ECM)
|
Input Power Supply (volts)
|
4.75 to 5.25 VDC
(Optional voltage regulator available for 6 to 32 VDC)
|
Power Consumption (current)
|
~40 mA typical @ 5 VDC
|
Calibration
|
None. Preconfigured for container materials or wall thickness.
|
Output Configuration
|
Open Collector, 40 mA max @ 30 VDC closed (Factory configurable for Wet or Dry sink)
|
Switch Condition
|
Normally Open/Normally Closed
|
Response Time
|
2 msec.
|
Low Noise (RFI/EMI)
|
10V per Meter Field Strength
|
Agency Approvals
|
UL 508 Listed (File E 305671), CE & IEC 61326 (RFI/EMI)
|
Temperature (sensor)*
|
32ºF to 158ºF (0ºC to 70ºC)
|
Mean Time Between Failure
|
50K Hours (Target)
|
Repeatability
|
±0.039" (±1 mm)
|
Accuracy
|
±0.063" (±1.6 mm)
|
Sealing Capability
|
IP 65
|
*Electronics are rated at [32°F to 149°F (0ºC to 65ºC)]
|
|
|