| Specifications |
Input
|
Pressure Range
|
0 to 415" wc (0 to 15 psi)
|
Proof Pressure
|
30 psi (<=80"wc) 60 psi (<=150"wc) 100 psi (>150"wc)
|
Burst Pressure
|
45 psi (<=28"wc) 60 psi (>28"wc to 80"wc) 90 psi (<=150"wc) 145 psi (>150"wc)
|
Fatigue Life
|
10 million FS cycles
|
Performance
|
Long Term Stability
|
0.25% span/annum
|
Accuracy
|
0.2% span max
|
Thermal Error
|
2% span max
|
Compensated Temperatures
|
-4°F to 140°F (-20°C to 60°C)
|
Operating Temperatures
|
-15°F to 185°F (-25°C to 85°C) Electrical Code G and L -5°F to 120°F (-20°C to 50°C) Electrical Code M and 3 -40°F to 212°F (-40°C to 100°C) Process Media
|
Zero Tolerance
|
0.1% span
|
Span Tolerance
|
0.1% span
|
Mounting Effects
|
0.25% span max
|
Response Time
|
5 ms
|
Supply Voltage Sensitivity
|
0.01% span/volt
|
Zero Adjustment
|
+/- 10% (by potentiometer)
|
Span Adjustment
|
+/- 10% (by potentiometer)
|
Mechanical Configuration
|
Pressure Port
|
See ordering guide
|
Wetted Parts
|
318 Duplex SS, Ceramic, Viton (Nitrile Optional)
|
Electrical Connection
|
See ordering guide
|
Enclosure
|
Code M and Code 3 IP68 Submersible Code G IP65
|
Vibration
|
35g peak sinusoidal, 5 to 2000 Hz, MIL STD 810, Method 514.2, Procedure I
|
Acceleration
|
100g, MIL STD 810C, Method 513.2, Procedure II
|
Approvals
|
CE ExII, 1G, EExia IIB Lloyds Register
|
Weight
|
330gms (excluding cable) (12oz)
|
Voltage Output Units
|
Output
|
See ordering guide (3-wire)
|
Supply Voltage (Vs)
|
9 to 35vdc (8-35vdc, 1-6vdc Output)
|
Current Output Units
|
Output
|
4-20 mA (2 wire)
|
Supply Voltage (Vs)
|
9 to 35 Vdc
|
Max Loop Resistance
|
(Vs-9) x 50 ohms
|